| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
University of Wisconsin Osteoporosis Clinical Center & Research Program, University of Wisconsin, Madison, Wisconsin 53705
Address all correspondence and requests for reprints to: Neil Binkley, M.D., 2870 University Avenue, Suite 100, Madison, Wisconsin 53705. E-mail: nbinkley{at}wisc.edu.
| Abstract |
|---|
|
|
|---|
Objectives: We investigated the current agreement of 25(OH)D measurement in clinical laboratories and explored the possibility that simple calibration would improve between-laboratory agreement.
Design and Participants: Serum obtained from healthy volunteers (age 20–60 yr) and one "calibrator," selected to have a 25(OH)D value near 30 ng/ml, were sent for 25(OH)D measurement in four clinical laboratories (laboratories A–D) using HPLC, liquid chromatography tandem mass spectroscopy, and RIA methodologies.
Main Outcome Measures: Serum 25(OH)D. Based upon self-report, the laboratory with the lowest interassay percent coefficient of variation was assigned as the reference to which the others were compared using linear regression and Bland-Altman analyses (Analyse-it; Analyse-it Software, Ltd., Leeds, UK).
Results: Good correlation was observed for 25(OH)D measurement between laboratory A and laboratories B–D (R2 = 0.99, 0.81, and 0.95, respectively). Modest between-laboratory variation was noted; the mean bias ranged from 2.9–5.2 ng/ml. Consistent with a systematic offset, each value in laboratory B was higher than in laboratory A, and 89% of values from laboratories B–D were higher than laboratory A. The use of a single calibrator and correction factor reduced mean between-laboratory bias for laboratories B and D.
Conclusions: Measurement of 25(OH)D by clinical laboratories yields similar results. The use of even a single calibrator will improve, but not resolve, between-laboratory variability. Based upon these data, in combination with reported within-individual variability, we recommend that clinicians aim for values greater than 30 ng/ml in their patients.
| Introduction |
|---|
|
|
|---|
| Subjects and Methods |
|---|
|
|
|---|
As part of quality control measures, over 200 laboratories worldwide participate in the DEQAS. The overall aim of this voluntary effort is to ensure analytical reliability of vitamin D assays. This is achieved through quarterly distribution of serum pools (n = 5 specimens) with subsequent reporting of these results to participants. Various methodologies are used for 25(OH)D measurement by laboratories participating in the DEQAS, including RIA, HPLC, and LCMSMS. DEQAS data are reported to participating laboratories in aggregate form with mean values reported. However, individual laboratory values are reported anonymously for HPLC and LCMSMS methods. The example data presented here are from the April, 2006, quarterly specimen distribution.
Study participants/clinical 25(OH)D measurement
Blood was obtained from 15 generally healthy adults (eight females, seven males; age 20–60 yr) by routine venipuncture, allowed to clot for 30 min at room temperature, then centrifuged. Study participants had no known acute or life-threatening illnesses and chemistry panels without clinically significant abnormalities. Serum aliquots were prepared and frozen at –80 C until shipped in a routine clinical manner for analysis. Aliquots of serum from one individual were obtained to serve as a "calibrator" that was known to contain approximately 30 ng/ml 25(OH)D from an independent analysis by HPLC. Aliquots of serum from all volunteers and the calibrator were sent as frozen clinical specimens to four laboratories (laboratories A–D) for 25(OH)D measurement. The methodologies used for 25(OH)D measurement included HPLC (laboratory A), LCMSMS (laboratories B and C), and RIA (DiaSorin, Inc., Stillwater, MN; laboratory D). Interassay 25(OH)D measurement precision [percent coefficient of variation (CV)] was obtained by self-report from each of the four laboratories (Table 1
). This work was approved by the University of Wisconsin Health Sciences institutional review board; informed consent was obtained from all study participants.
|
Based upon lowest self-reported percent CV, serum 25(OH)D results from laboratory A were defined for the purposes of this report as the "gold standard" to which the other laboratories were compared. Clinical vitamin D status was defined by serum 25(OH)D measurement as follows: optimal vitamin D status, more than or equal to 30 ng/nl; low vitamin D status, less than 30 ng/ml. Serum 25(OH)D measurements from the four laboratories were compared using linear regression and Bland-Altman analyses (Analyse-it; Analyse-it Software, Ltd., Leeds, UK), an accepted methodology to compare two measurement techniques (5). Systematic bias was assessed by determination of the difference of each measurement in laboratories B–D compared with laboratory A ["delta" 25(OH)D]. For data "correction," each individual value from laboratories B–D was subsequently multiplied by the ratio of the "calibrator" value obtained in each respective laboratory compared with that obtained in laboratory A. After application of this correction factor to each individual value, 25(OH)D results from laboratories B–D were again compared with those of laboratory A using linear regression and Bland-Altman analyses as noted previously.
| Results |
|---|
|
|
|---|
Results are arbitrarily presented for two serum specimens (defined here as "1" and "2"). For these two sera, good agreement was reported for the mean 25(OH)D value from laboratories using HPLC (n = 9), LCMSMS (n = 8), and the DiaSorin RIA (n = 54) methodologies. Specifically, the mean (SD) values reported for these three technologies were 35.7 (8.4), 34.6 (7.3), and 31.4 (6.4) ng/ml for serum 1, and 25.5 (7.1), 28.9 (5.0), and 26.4 (5.6) ng/ml for serum 2, respectively.
The DEQAS reports individual values for those laboratories using HPLC or LCMSMS, which upon review appear to demonstrate substantial between-laboratory variability (Fig. 1A
). Evaluation of 25(OH)D results for these two sera by individual laboratory using HPLC methodology identifies what may be a systematic offset that suggests the possibility that assay calibration differences might explain some of the between-laboratory differences (Fig. 1B
).
|
Of these 15 specimens, 10 contained only 25(OH)D3 as identified by HPLC. In the remaining five specimens, the 25(OH)D2 concentration ranged from 5–19 ng/ml. Because clinical decisions are made based upon the total 25(OH)D concentration, this value was used for between-laboratory comparison. Diagnostic categorization (optimal vs. low vitamin D status) was identical in 80% (12 of 15) of subjects (Fig. 2
). Good correlation was observed between laboratory A and the other three laboratories (R2 = 0.99, 0.81, and 0.95) (Fig. 3
, A–C). However, modest between-laboratory variability was noted; the mean bias of laboratories B–D ranged from +2.9 to +5.1 ng/ml when compared with laboratory A (example in Fig. 3D
). This systematic positive bias leads to 89% (40 of 45) of values being higher in laboratories B–D than the corresponding laboratory A value (Fig. 4
). "Correction" of the 25(OH)D values obtained by laboratories B–D reduces the mean bias (Fig. 5
, A–C), e.g. after correction, laboratories A and B obtain virtually identical results (Fig. 5
, A and D).
|
|
|
|
| Discussion |
|---|
|
|
|---|
There has been considerable discussion of what constitutes optimal vitamin D status and what cutpoint to use for a clinical diagnosis of vitamin D "deficiency" or "inadequacy." (7, 8, 9) As could be expected, the prevalence of low D status is highly dependent upon the cutpoint selected (10). Nonetheless, there is increasing clinical consensus that values less than 30–32 ng/ml are indicative of suboptimal vitamin D status (7). It is important for clinicians to recognize that any such arbitrary cutpoint will have variability surrounding the value chosen from both analytical and biological sources. Specifically, analytical imprecision is present in all quantitative medical procedures due to human and instrument limitations. In addition, at this point in time, there is no internationally recognized primary standard, however, the National Institute of Standards and Technology is currently working to rectify this situation. Based upon these results, availability of standards should enhance between-laboratory agreement.
In addition to analytical variability, within-individual biological variability of the parameter being measured, in this case 25(OH)D, must also be considered. Unfortunately, this aspect of 25(OH)D measurement has received only limited investigation. One study found 13–19% within-individual variability in 25(OH)D (11), which caused the authors to question the clinical use of a single 25(OH)D measurement in an individual patient. Although it is possible that this report reflects the well-known seasonal variability of vitamin D status due to variations in sunlight exposure (12, 13, 14, 15), scrutiny of the current clinical approach of diagnosing vitamin D status based upon a single 25(OH)D measurement is clearly warranted. Furthermore, the study of within-individual variability of 25(OH)D in people on a stable vitamin D intake is clearly warranted. Moreover, for those clinicians using infrequent dosing of large amounts of vitamin D, consideration of measuring trough concentration appears to be necessary because 25(OH)D levels peak within 1 wk after dosing with ergocalciferol (D2) and subsequently decline (16). Preliminary data suggest that the measurement of trough 25(OH)D concentration may be appropriate when dosing with monthly D3 as well (17). Additional study of this phenomenon is warranted.
Although further work in improving analytical 25(OH)D assay performance and investigating biological variability is indicated, clinicians are currently measuring 25(OH)D, making diagnostic assessments, and providing treatment recommendations based on clinical 25(OH)D measurement. Due to the assay variability reported here, and current (albeit limited) biological variability data, we believe it appropriate that clinicians aim to achieve 25(OH)D values above 30 ng/ml. Some clinicians may wish to consider aiming for a 25(OH)D concentration of approximately 40 ng/ml. This suggestion is based upon the increasingly accepted recommendation that a 25(OH)D concentration of more than or equal to 30 ng/ml be considered optimal, combined with the reported approximate 20% within-individual variability (i.e. 6 ng/ml if one targets 30 ng/ml) and the approximate 5 ng/ml systematic bias reported here. We recognize that some individuals with a serum 25(OH)D value of 40 ng/ml could truly have a value greater than this, but given the wide margin of safety reported for vitamin D, such a recommendation seems clinically prudent to ensure that the vitamin D status of an individual is optimal (>30 ng/ml).
Limitations of this study include the small sample size, limited population studied, and the fact that not all methodologies available for 25(OH)D measurement were evaluated. In this regard, it was our intent to compare the performance of 25(OH)D assays clinically available to us; we did not seek to evaluate all available assays or results performed in research settings. However, with the interest in automated technology, we did evaluate one such instrument, the DiaSorin Liaison in a research environment. In this setting, we found good correlation (R2 = 0.95) with laboratory A but feel it not justifiable to include as primary data because we did not have access to this technology in a routine clinical setting. Finally, we recognize that it is possible that other technologies and the study of other populations would provide differing results than reported here.
In conclusion, there is good agreement of serum 25(OH)D measured in these four clinical laboratories. Between-laboratory agreement should be improved with the use of standard calibrators. Based upon analytical and biological variability, it seems prudent that clinicians target a serum 25(OH)D value greater than 30 ng/ml to ensure vitamin D adequacy in their patients.
| Footnotes |
|---|
First Published Online March 4, 2008
Abbreviations: CV, Coefficient of variation; DEQAS, vitamin D External Quality Assessment Scheme; LCMSMS, liquid chromatography tandem mass spectroscopy; 25(OH)D, 25-hydroxy-vitamin D.
Received October 22, 2007.
Accepted February 25, 2008.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
R. Eastell, A. Arnold, M. L. Brandi, E. M. Brown, P. D'Amour, D. A. Hanley, D. S. Rao, M. R. Rubin, D. Goltzman, S. J. Silverberg, et al. Diagnosis of Asymptomatic Primary Hyperparathyroidism: Proceedings of the Third International Workshop J. Clin. Endocrinol. Metab., February 1, 2009; 94(2): 340 - 350. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. W Norman A vitamin D nutritional cornucopia: new insights concerning the serum 25-hydroxyvitamin D status of the US population Am. J. Clinical Nutrition, December 1, 2008; 88(6): 1455 - 1456. [Full Text] [PDF] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| Endocrinology | Endocrine Reviews | J. Clin. End. & Metab. |
| Molecular Endocrinology | Recent Prog. Horm. Res. | All Endocrine Journals |